Z-1901-2019Class II

Siemens Artis Q zeego ---Interventional Fluoroscopic X-Ray System Model Number: 10848283

Siemens Medical Solutions USA, Inc. / initiated 2019-06-06 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Siemens Artis Q zeego ---Interventional Fluoroscopic X-Ray System Model Number: 10848283 on . FDA classified it as Class II and gives the reason as "Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1901-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Artis Q zeego ---Interventional Fluoroscopic X-Ray System Model Number: 10848283
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel
Root cause tag
design
FDA root cause
Device Design
Action
Siemens issued letter dated 6/6/19 advising of the problem, health risk and action to take: .Siemens is implementing a corrective action to all potentially affected users via Update Instruction AX023/18/S. Siemens Service organization will replace the screwed connection of the ceiling support according to specifications / instructions. In the event the bolt connection cannot be replaced due to technical conditions, the complete bearing will be replaced.
Quantity in commerce
2
Distribution
Nationwide
Product codes and lots
Serial Numbers: 117309 117125

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1901-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1894-2019Class IITerminatedVoluntary: Firm initiated
Z-1895-2019Class IITerminatedVoluntary: Firm initiated
Z-1896-2019Class IITerminatedVoluntary: Firm initiated
Z-1897-2019Class IITerminatedVoluntary: Firm initiated
Z-1898-2019Class IITerminatedVoluntary: Firm initiated
Z-1899-2019Class IITerminatedVoluntary: Firm initiated
Z-1900-2019Class IITerminatedVoluntary: Firm initiated
Z-1901-2019Class IITerminatedVoluntary: Firm initiated
Z-1902-2019Class IITerminatedVoluntary: Firm initiated
Z-1903-2019Class IITerminatedVoluntary: Firm initiated
Z-1904-2019Class IITerminatedVoluntary: Firm initiated
Z-1905-2019Class IITerminatedVoluntary: Firm initiated