Z-1963-2019Class II
Ultra-Drive Irrigation Tubing Assembly, Reference Number 423834 -
Zimmer GmbH / initiated 2019-06-04 / Terminated / root cause: labeling and UDI
Zimmer Biomet, Inc. began a recall of Ultra-Drive Irrigation Tubing Assembly, Reference Number 423834 - Product Usage: Item used with the UltraDrive System to provide proper irrigation during the... on . FDA classified it as Class II and gives the reason as "The products do not have sufficient data to support the labeled shelf life of 10 years". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1963-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Ultra-Drive Irrigation Tubing Assembly, Reference Number 423834 - Product Usage: Item used with the UltraDrive System to provide proper irrigation during the cutting and removal of bone and acrylic bone cement in orthopedic applications.
- Product code
- JDX Instrument, Surgical, Sonic And Accessory/Attachment
- Reason for recall
- The products do not have sufficient data to support the labeled shelf life of 10 years.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- Urgent Medical Device Recall notification letters dated 6/4/19 were sent to customers.
- Quantity in commerce
- 8478 total
- Distribution
- Worldwide distribution - US nationwide distribution and countries of Australia, Brazil, Canada, Chile, China, Columbia, Hong Kong, India, Japan, Malaysia, Netherlands, and Taiwan.
- Product codes and lots
- Lot Numbers: 211325 209586 210002 210488 210566 210848 210867 210972 211172 211237 212076 212488 212622 212928 212966 214005 214051 214118 214255 214499 214546 214669 214822 214995 215048 215329 215445 215524 215623 216489 216611 217370 217556 217761 217767 218667 219305 219661 219913 220935
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1962-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1963-2019 | Class II | Terminated | Voluntary: Firm initiated |