ColoCARE, Cat. No. 5650, packaged 250 envelopes/box, For In-Vitro Diagnostic Use. The firm name on the label...
Helena Laboratories, Corp. began a recall of ColoCARE, Cat. No. 5650, packaged 250 envelopes/box, For In-Vitro Diagnostic Use. The firm name on the label is Helena Laboratories, Beaumont, TX on . FDA classified it as Class II and gives the reason as "The positive control on some tests of the lots did not appropriately react within the specified timeframe to verify the test was functioning correctly". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1949-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Helena Laboratories, Corp.
- Product
- ColoCARE, Cat. No. 5650, packaged 250 envelopes/box, For In-Vitro Diagnostic Use. The firm name on the label is Helena Laboratories, Beaumont, TX.
- Product code
- KHE Reagent, Occult Blood
- Reason for recall
- The positive control on some tests of the lots did not appropriately react within the specified timeframe to verify the test was functioning correctly.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The recalling firm made notifications beginning 8/8/2018 via phone and letter delivered via email. Customers were to dispose of the product.
- Quantity in commerce
- 27/250 pouch kits
- Distribution
- Distribution was made to AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MO, MS, MT, NC, ND, NE, NV, NY, OH, OK, PA, SD, TN, TX, VA, WA, and WY. There was no foreign/military/government distribution.
- Product codes and lots
- Kit lots 1-17-5650, exp. 8/31/2018, and 2-17-5650, exp. 9/30/2018. These kit lots contain ColoCARE pad lot numbers 1-17-551039, exp. 8/31/2018, and 2-17-551039, exp. 9/30/2018.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1949-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1950-2019 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.