Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1920-2019Class II

SOMATOM go.Now, Model No. 11061618 with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1

Siemens Medical Solutions USA, Inc. / initiated 2019-06-04 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of SOMATOM go.Now, Model No. 11061618 with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to... on . FDA classified it as Class II and gives the reason as "The potential sporadic performance problems may cause scanning workflow interruptions and unexpected user notifications resulting in diagnostic delay or need for patient rescan". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1920-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM go.Now, Model No. 11061618 with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. This CT system can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The potential sporadic performance problems may cause scanning workflow interruptions and unexpected user notifications resulting in diagnostic delay or need for patient rescan.
Root cause tag
software
FDA root cause
Software design
Action
On June 4, 2019, the firm notified customers of the recall via Urgent Medical Device Correction letters. Customers were informed that the firm has developed software update CT VA20A-SP2A to ensure uninterrupted scanning workflows and to reduce the number of user notifications. This update will also improve some areas during the start-up and shut-down processes, as well as providing other improvements related to the system, examination and reconstruction processes, and dual energy and intervention . The corrective action will be provided free of charge and will be distributed via Update Instruction CT026/19P. This update will be performed remotely. If you have any questions, please contact our service organization at 1- 800-888-7436. Customers were asked to promptly notify and instruct all staff at their organizations who need to be aware of this notice. In addition, if you have further distributed this product, please identify your customers and notify them at once of this product advisory notice. If the device has been sold and is therefore no longer in your possession, please forward this notice to the new owner.
Quantity in commerce
69
Distribution
US Nationwide distribution in the states of AL CA CT FL GA IL IN KS KY LA ME MI MO MS NC ND NE NH NY OH OK SD TN TX VA and WI
Product codes and lots
106527 106565 106456 106464 106545 106237 106375 106526 106484 106562 106400 106260 106459 106150 106593 106133 106115 106360

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1920-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1919-2019Class IITerminatedVoluntary: Firm initiated
Z-1920-2019Class IITerminatedVoluntary: Firm initiated
Z-1921-2019Class IITerminatedVoluntary: Firm initiated
Z-1922-2019Class IITerminatedVoluntary: Firm initiated