RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayPlan 1...
RAYSEARCH LABORATORIES AB began a recall of RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayPlan 1 (RayStation 4.9) and RayPlan 2 (RayStation 6) . Intended... on . FDA classified it as Class II and gives the reason as "When calculating electron Monte Carlo dose with a very large number of histories, the dose calculation may be wrong". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1930-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RaySearch Laboratories AB (publ)
- Product
- RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayPlan 1 (RayStation 4.9) and RayPlan 2 (RayStation 6) . Intended Use: software system designed for treatment planning and analysis of radiation therapy
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- When calculating electron Monte Carlo dose with a very large number of histories, the dose calculation may be wrong
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- RaySearch issued Field Safety Notice, Medical Device Correction #24138 via email on November 20th, 2017. The letter identifies the product, issue, health risk and action to take: Do not use a number of histories above 6 million for electron dose calculation. Inspect all commissioned electron beam models for any applicator calculated with greater than 6 million. Contact customer support if you have any further questions. Confirm you have read and understood this notice. This issue will be resolved in the next version of RayStation, scheduled for market release in December 2017. Educate planning staff and all users about this workaround.
- Quantity in commerce
- 744
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Software versions: 3.5.0.16 , 3.5.1.6 , 4.0.0.14 , 4.0.1.4 , 4.0.2.9 , 4.0.3.4 , 4.5.0.19 , 4.5.1.14 , 4.5.2.7 , 4.7.0.15 , 4.7.1.10 , 4.7.2.5 , 4.7.3.13 , 4.7.4.4 , 4.7.5.4 , 4.9.0.42 , 5.0.0.37 , 5.0.1.11 , 5.0.2.35 , 5.0.3.17 , 6.0.0.24 , 6.1.0.26 , 6.1.1.2 and 6.2.0.7
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1930-2019 | Class II | Terminated | Voluntary: Firm initiated |