Z-2150-2019Class II

Human Assayed Multi-Sera Level 2, Model NO. HN1530

Randox Laboratories, Limited / initiated 2019-06-11 / Terminated

Randox Laboratories, Limited began a recall of Human Assayed Multi-Sera Level 2, Model NO. HN1530 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of diagnostic... on . FDA classified it as Class II and gives the reason as "The RX Series control target and range value for ALT (Tris buffer without P5P 37C), Bicarbonate and Gamma-GT has been incorrectly assigned in specific lots of the Human Assayed Multi-Sera Controls". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2150-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Randox Laboratories, Limited
Product
Human Assayed Multi-Sera Level 2, Model NO. HN1530 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of diagnostic assays. The Human Assayed Multi-sera is for the control of accuracy.
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
The RX Series control target and range value for ALT (Tris buffer without P5P 37C), Bicarbonate and Gamma-GT has been incorrectly assigned in specific lots of the Human Assayed Multi-Sera Controls.
FDA root cause
Under Investigation by firm
Action
On May 25, 2019, the firm notified its customers of the recall via Urgent Medical Device Correction letters. Customers were instructed to do the following: " Review your reagent inventory of these products. Remove all previous versions of the Value Sheets in use and replace with the revised version. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to t echnical.services@ randox.com within five working days. If you have any questions or concerns, please contact Randox Technical Services.
Quantity in commerce
149 (US)
Distribution
US Nationwide Distribution - AZ CA CT FL GA IL KY LA MA MD MI MN MO NJ NY OH OK OR PA TX UT VA WA WI WV and PR. Worldwide foreign distribution.
Product codes and lots
1306UN, 1308UN, 1309UN

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2150-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2150-2019Class IITerminatedVoluntary: Firm initiated