1 pc Full Wrap Support Belt Apron, Product Number F111, Part Numbers MTREF111*F, MTREF111E, MTREF111F...
Burlington Medical, LLC began a recall of 1 pc Full Wrap Support Belt Apron, Product Number F111, Part Numbers MTREF111*F, MTREF111E, MTREF111F, REF111*F, REF111*M, REF111B, REF111E, REF111F, SEF111*F... on . FDA classified it as Class II and gives the reason as "The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2111-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Burlington Medical LLC
- Product
- 1 pc Full Wrap Support Belt Apron, Product Number F111, Part Numbers MTREF111*F, MTREF111E, MTREF111F, REF111*F, REF111*M, REF111B, REF111E, REF111F, SEF111*F, SEF111*M, SEF111C, SEF111D, SEF111E, SEF111F, UEF111*F, UEF111*M, UEF111A, UEF111B, UEF111C, UEF111D, UEF111E, UEF111XLM, UEF111XXLM Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.
- Product code
- EAJ Apron, Leaded
- Reason for recall
- The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- Urgent Medical Device Recall notification letters were sent to customers on 6/7/19. Please take the following actions: Cease using recalled products immediately, quarantine the products, and prepare the products for return shipment to Burlington Medical. Contact Burlington Medical at (800) 221-3466 for specific return instructions. Note that Burlington Medical Customer Care will work with you to facilitate the return of recalled products, including by issuing you a Recall Tracking number. Burlington Medical will replace all recalled products at no cost to you. The replacement product will be the same size and model as the recalled product; if the same size and model is not available or has been discontinued, then Burlington Medical will replace the recalled product with a Burlington Medical product that is similar to the recalled product. Our Customer Care team will arrange for free shipment of the replacement product. If you have any questions, please call your Burlington Medical representative or Burlington Medical Customer Care at (800) 221-3466. The recall was expanded an additional Urgent Medical Device Recall notification letters were sent to customers on 7/3/19.
- Quantity in commerce
- 51
- Distribution
- US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.
- Product codes and lots
- Serial Numbers: 379682 379535 423744 423745 387796 426462 419802 384492 377568 379681 379690 379698 386085 377724 382991 384347 385884 386083 381755 379715 385640 381868 426861 426802 426955 426895 426634 426809 426951 416567 426461 426789 426952 381869 443767 443802 377712 381865 377711 382966 381899 381866 382876 387805 382949 385982 423502 339345 423501 414165 414167
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2107-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2108-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2109-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2110-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2111-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2112-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2113-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2114-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2115-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2116-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2117-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2118-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2119-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2120-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2121-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2122-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2123-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2124-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2125-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2126-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2127-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2128-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2129-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2130-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.