The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent...
Beckman Coulter Inc. began a recall of The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a... on . FDA classified it as Class II and gives the reason as "Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2763-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.
- Product code
- CGR Radioimmunoassay, Cortisol
- Reason for recall
- Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- An Urgent Medical Device Recall Letter was sent during the week of 9/17/2018 to all affected customers via mail and or email. Customers were instructed to discontinue use and discard remaining product. Customers were also instructed to consider a review of the Cortisol results generated with the affected lot number at the discretion of the Medical Director.
- Quantity in commerce
- Access Cortisol Reagent Assay
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Lithuania, Mexico, Panama, Serbia, South Africa, and Vietnam.
- Product codes and lots
- Lot Number 831755.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2763-2019 | Class II | Terminated | Voluntary: Firm initiated |