Z-2151-2019Class II

Liberty Cycler Cassettes (a disposable cassette that is intended to be used to perform peritoneal dialysis...

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2019-05-29 / Terminated

Fresenius Medical Care Renal Therapies Group, LLC began a recall of Liberty Cycler Cassettes (a disposable cassette that is intended to be used to perform peritoneal dialysis (APD) in conjunction with the Liberty cycler as... on . FDA classified it as Class II and gives the reason as "Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler and is a potential risk for...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2151-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Liberty Cycler Cassettes (a disposable cassette that is intended to be used to perform peritoneal dialysis (APD) in conjunction with the Liberty cycler as follows: (1) LIBERTY Line Cycler Set with Stay-safe connectors (Dual Patient Connector) Part Number: 050-87212 (2) LIBERTY Line Cycler Set with Stay-safe connectors (Single Patient Connector) Part Number: 050-87215 (3) LIBERTY Integrated Cycler Set with single Stay-safe patient connector and extended drain line Part Number: 050-87216
Product code
FKX System, Peritoneal, Automatic Delivery
Reason for recall
Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler and is a potential risk for contamination.
FDA root cause
Use error
Action
Fresenius issued on 05/29/19 The IMPORTANT MEDICAL DEVICE INFORMATION notification explains to the Home Therapy Nurses/Clinics to review the new flyer with each of their APD patients who use a Liberty Cycler. For the patient notification, it will encourage them to use and follow the attached instructional flyer during set up for their daily dialysis treatment. The notification will have a copy of the new step-by-step instructional flyer. Included is a Reply Form for confirmation the consignee read and understands the notice sent to them.
Quantity in commerce
33732973 Cassettes; 18,674 Liberty Cyclers
Distribution
US Nationwide Distribution
Product codes and lots
All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2151-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2151-2019Class IITerminatedVoluntary: Firm initiated