Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2083-2019Class II

RX Daytona, Model Nos. RX4040 and RX4041

Randox Laboratories, Limited / initiated 2019-06-24 / Terminated

Randox Laboratories, Limited began a recall of RX Daytona, Model Nos. RX4040 and RX4041 on . FDA classified it as Class II and gives the reason as "There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2083-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Randox Laboratories, Limited
Product
RX Daytona, Model Nos. RX4040 and RX4041
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.
FDA root cause
Under Investigation by firm
Action
On June 24, 2019, the firm notified its customers of the recall through an Urgent Field Safety Notice. Customers were asked to take the following actions: If there is any sign of damage to the RCU lid or lid sensor " Switch of main analyser main power supply as soon as possible " Contact technical support immediately. " Discuss the contents of this notice with your Medical Director. " Display this notice in a prominent location near the instrument. " Complete and return the response form 12187-QA to [email protected] within five working days.
Quantity in commerce
3
Distribution
Domestic distribution to CA CT MA MD MO NC NJ NY OH PA UT WV and PR. Worldwide foreign distribution.
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2083-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2083-2019Class IITerminatedVoluntary: Firm initiated
Z-2084-2019Class IITerminatedVoluntary: Firm initiated