Z-2084-2019Class II
RX Imola, Model Nos. RX4900
Randox Laboratories, Limited / initiated 2019-06-24 / Terminated
Randox Laboratories, Limited began a recall of RX Imola, Model Nos. RX4900 on . FDA classified it as Class II and gives the reason as "There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2084-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Randox Laboratories, Limited
- Product
- RX Imola, Model Nos. RX4900
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.
- FDA root cause
- Under Investigation by firm
- Action
- On June 24, 2019, the firm notified its customers of the recall through an Urgent Field Safety Notice. Customers were asked to take the following actions: If there is any sign of damage to the RCU lid or lid sensor " Switch of main analyser main power supply as soon as possible " Contact technical support immediately. " Discuss the contents of this notice with your Medical Director. " Display this notice in a prominent location near the instrument. " Complete and return the response form 12187-QA to [email protected] within five working days.
- Quantity in commerce
- 27
- Distribution
- Domestic distribution to CA CT MA MD MO NC NJ NY OH PA UT WV and PR. Worldwide foreign distribution.
- Product codes and lots
- All serial numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2083-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2084-2019 | Class II | Terminated | Voluntary: Firm initiated |