Z-0030-2020Class II

L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene...

Baxter Healthcare Corporation / initiated 2018-10-10 / Terminated / root cause: sterility

Baxter Healthcare Corporation began a recall of L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene oxide on . FDA classified it as Class II and gives the reason as "Potential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-0030-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene oxide.
Product code
FKX System, Peritoneal, Automatic Delivery
Reason for recall
Potential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with micro-organisms. This may predispose the patient to peritonitis.
Root cause tag
sterility
FDA root cause
Mixed-up of materials/components
Action
On 10/10/2018, and 10/15/2018, Urgent Product Recall notices were mailed to dialysis providers at medical centers. Customers were asked to do the following: 1. Locate and remove the affected product lot from your facility. The product code and lot number can be found on the individual product or shipping carton. 2. Contact recalling firm for Service to arrange for return and credit. Center for Service can be reached at 888-229-0001. 3. If you purchased this product directly from the recalling firm, complete and return the enclosed customer reply form. 4. If you purchased this product from a distributor, please contact your distributor for return and credit. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a consumer-level recall of the affected product that you distributed to customers. Customers with additional questions are encouraged to contact: Corporate Product Surveillance at 800-437-5176, Clinical Helpline at 888-736-2543, and Renal Therapy Services at 800-553-6898. On 10/15/2018, Urgent Product Recall notices were mailed to Peritoneal Dialysis Patients who were asked to do the following: 1. Locate and remove the affected product lot from your supplies. The product code and lot number can be found on the individual product or shipping carton. 2. Contact recalling firm to arrange for return and credit. HomeCare Services can be reached at 800-284-4060. 3. Complete and return the enclosed Home Patient Reply Form.
Quantity in commerce
8640 eaches
Distribution
U.S.: VA, DC, CA, FL, SC, MD, DE, PA, NC, WA, GA, TX, NV. No foreign (OUS) distribution.
Product codes and lots
Lot Number H18D22030

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0030-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0030-2020Class IITerminatedVoluntary: Firm initiated