Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2308-2019Class II

BD Vacutainer LH (Lithium Heparin) 34 I.U. Plus Blood Collection Tubes, Catalog Number 368494

Becton Dickinson & Co / initiated 2019-06-12 / Terminated / root cause: design

Becton Dickinson & Company began a recall of BD Vacutainer LH (Lithium Heparin) 34 I.U. Plus Blood Collection Tubes, Catalog Number 368494 on . FDA classified it as Class II and gives the reason as "False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2308-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Vacutainer LH (Lithium Heparin) 34 I.U. Plus Blood Collection Tubes, Catalog Number 368494
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.
Root cause tag
design
FDA root cause
Device Design
Action
Urgent Medical Device Correction notification letters dated 6/12/19 were sent to customers.
Quantity in commerce
1,436,228,004 total
Distribution
Worldwide Distribution, including US Nationwide
Product codes and lots
UDI: (01)30382903680352

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2308-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2296-2019Class IITerminatedVoluntary: Firm initiated
Z-2297-2019Class IITerminatedVoluntary: Firm initiated
Z-2298-2019Class IITerminatedVoluntary: Firm initiated
Z-2299-2019Class IITerminatedVoluntary: Firm initiated
Z-2300-2019Class IITerminatedVoluntary: Firm initiated
Z-2301-2019Class IITerminatedVoluntary: Firm initiated
Z-2302-2019Class IITerminatedVoluntary: Firm initiated
Z-2303-2019Class IITerminatedVoluntary: Firm initiated
Z-2304-2019Class IITerminatedVoluntary: Firm initiated
Z-2305-2019Class IITerminatedVoluntary: Firm initiated
Z-2306-2019Class IITerminatedVoluntary: Firm initiated
Z-2307-2019Class IITerminatedVoluntary: Firm initiated
Z-2308-2019Class IITerminatedVoluntary: Firm initiated
Z-2309-2019Class IITerminatedVoluntary: Firm initiated
Z-2310-2019Class IITerminatedVoluntary: Firm initiated