Z-2012-2019Class II

AXIOM Artis FC, Model Number 5904433

Siemens Medical Solutions USA, Inc. / initiated 2019-06-12 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of AXIOM Artis FC, Model Number 5904433 on . FDA classified it as Class II and gives the reason as "A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2012-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
AXIOM Artis FC, Model Number 5904433
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
Root cause tag
design
FDA root cause
Component design/selection
Action
Urgent Medical Device Correction notification letters dated 6/12/19 were sent to customers. In the event the issue occurs, the system cannot be operated normally as X-ray will not possible. Planned procedures may have to be terminated and performed on an alternative X-ray system. Siemens advises that an alternative system is available to continue treatment in case of occurrence. Siemens is implementing a corrective action to all potentially affected users of the Artis Zee and AXIOM Artis systems with generator A100 via Update Instruction AX038/18/S. The Siemens Service organization will modify the affected supply circuit and revise the wiring. This modification will bring the voltage back to the center of the tolerance range. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated and any recurrence of this potential fault is prevented.
Quantity in commerce
5
Distribution
US Nationwide Distribution
Product codes and lots
Serial Numbers: 10436 10444 10425 10433 10435

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2012-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2012-2019Class IITerminatedVoluntary: Firm initiated
Z-2013-2019Class IITerminatedVoluntary: Firm initiated
Z-2014-2019Class IITerminatedVoluntary: Firm initiated
Z-2015-2019Class IITerminatedVoluntary: Firm initiated
Z-2016-2019Class IITerminatedVoluntary: Firm initiated
Z-2017-2019Class IITerminatedVoluntary: Firm initiated
Z-2018-2019Class IITerminatedVoluntary: Firm initiated
Z-2019-2019Class IITerminatedVoluntary: Firm initiated
Z-2020-2019Class IITerminatedVoluntary: Firm initiated
Z-2021-2019Class IITerminatedVoluntary: Firm initiated
Z-2022-2019Class IITerminatedVoluntary: Firm initiated
Z-2023-2019Class IITerminatedVoluntary: Firm initiated
Z-2024-2019Class IITerminatedVoluntary: Firm initiated
Z-2025-2019Class IITerminatedVoluntary: Firm initiated
Z-2026-2019Class IITerminatedVoluntary: Firm initiated
Z-2027-2019Class IITerminatedVoluntary: Firm initiated
Z-2028-2019Class IITerminatedVoluntary: Firm initiated
Z-2029-2019Class IITerminatedVoluntary: Firm initiated
Z-2030-2019Class IITerminatedVoluntary: Firm initiated
Z-2031-2019Class IITerminatedVoluntary: Firm initiated
Z-2032-2019Class IITerminatedVoluntary: Firm initiated
Z-2033-2019Class IITerminatedVoluntary: Firm initiated
Z-2034-2019Class IITerminatedVoluntary: Firm initiated