Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2023-2019Class II

AXIOM Artis dMP, Model Number 7555365

Siemens Medical Solutions USA, Inc. / initiated 2019-06-12 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of AXIOM Artis dMP, Model Number 7555365 on . FDA classified it as Class II and gives the reason as "A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2023-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
AXIOM Artis dMP, Model Number 7555365
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
Root cause tag
design
FDA root cause
Component design/selection
Action
Urgent Medical Device Correction notification letters dated 6/12/19 were sent to customers. In the event the issue occurs, the system cannot be operated normally as X-ray will not possible. Planned procedures may have to be terminated and performed on an alternative X-ray system. Siemens advises that an alternative system is available to continue treatment in case of occurrence. Siemens is implementing a corrective action to all potentially affected users of the Artis Zee and AXIOM Artis systems with generator A100 via Update Instruction AX038/18/S. The Siemens Service organization will modify the affected supply circuit and revise the wiring. This modification will bring the voltage back to the center of the tolerance range. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated and any recurrence of this potential fault is prevented.
Quantity in commerce
32
Distribution
US Nationwide Distribution
Product codes and lots
Serial Numbers: 57204 57089 57077 57024 57144 57189 57208 57035 57188 57073 57072 57016 57009 57050 57207 57111 57171 57113 57193 57008 57198 57155 57124 57004 57105 57098 57116 57045 57156 58701 57183 57187

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2023-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2012-2019Class IITerminatedVoluntary: Firm initiated
Z-2013-2019Class IITerminatedVoluntary: Firm initiated
Z-2014-2019Class IITerminatedVoluntary: Firm initiated
Z-2015-2019Class IITerminatedVoluntary: Firm initiated
Z-2016-2019Class IITerminatedVoluntary: Firm initiated
Z-2017-2019Class IITerminatedVoluntary: Firm initiated
Z-2018-2019Class IITerminatedVoluntary: Firm initiated
Z-2019-2019Class IITerminatedVoluntary: Firm initiated
Z-2020-2019Class IITerminatedVoluntary: Firm initiated
Z-2021-2019Class IITerminatedVoluntary: Firm initiated
Z-2022-2019Class IITerminatedVoluntary: Firm initiated
Z-2023-2019Class IITerminatedVoluntary: Firm initiated
Z-2024-2019Class IITerminatedVoluntary: Firm initiated
Z-2025-2019Class IITerminatedVoluntary: Firm initiated
Z-2026-2019Class IITerminatedVoluntary: Firm initiated
Z-2027-2019Class IITerminatedVoluntary: Firm initiated
Z-2028-2019Class IITerminatedVoluntary: Firm initiated
Z-2029-2019Class IITerminatedVoluntary: Firm initiated
Z-2030-2019Class IITerminatedVoluntary: Firm initiated
Z-2031-2019Class IITerminatedVoluntary: Firm initiated
Z-2032-2019Class IITerminatedVoluntary: Firm initiated
Z-2033-2019Class IITerminatedVoluntary: Firm initiated
Z-2034-2019Class IITerminatedVoluntary: Firm initiated