Artis zee biplane MN, Model Number 10094143
Siemens Medical Solutions USA, Inc began a recall of Artis zee biplane MN, Model Number 10094143 on . FDA classified it as Class II and gives the reason as "A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2033-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Artis zee biplane MN, Model Number 10094143
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Urgent Medical Device Correction notification letters dated 6/12/19 were sent to customers. In the event the issue occurs, the system cannot be operated normally as X-ray will not possible. Planned procedures may have to be terminated and performed on an alternative X-ray system. Siemens advises that an alternative system is available to continue treatment in case of occurrence. Siemens is implementing a corrective action to all potentially affected users of the Artis Zee and AXIOM Artis systems with generator A100 via Update Instruction AX038/18/S. The Siemens Service organization will modify the affected supply circuit and revise the wiring. This modification will bring the voltage back to the center of the tolerance range. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated and any recurrence of this potential fault is prevented.
- Quantity in commerce
- 3
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Serial Numbers: 149202 149100 149106
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2012-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2013-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2014-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2015-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2016-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2017-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2018-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2019-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2020-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2021-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2022-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2023-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2024-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2025-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2026-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2027-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2028-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2029-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2030-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2031-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2032-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2033-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2034-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.