Z-2202-2019Class II

VITROS Chemistry Products GLU Slides-

Ortho-Clinical Diagnostics, Inc. / initiated 2019-06-17 / Terminated / root cause: manufacturing process

Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products GLU Slides-Product Code:1707801. IVD VITROS Chemistry Products GLU Slides quantitatively measure glucose (GLU) concentration in... on . FDA classified it as Class II and gives the reason as "Potentially Biased Results using VITROS Chemistry Products GLU Slides". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2202-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products GLU Slides-Product Code:1707801. IVD VITROS Chemistry Products GLU Slides quantitatively measure glucose (GLU) concentration in serum, plasma, urine, and cerebrospinal fluid using VITROS 250/350/5,1 FS/4600 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
Potentially Biased Results using VITROS Chemistry Products GLU Slides
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Ortho issued On 17 June 2019, a customer letter (Ref. CL2019-172) via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to customers and distributors. The customers were notified of the issue and instructed: Discontinue using Lot 0048-0964-9287, discard any remaining inventory and switch to an alternate lot. " If you do NOT have an alternate lot, consider the following alternate options: o Use an alternate glucose method until your replacement order arrives. o Consult with your medical director to determine if the bias observed and rate of occurrence when using the affected lot is acceptable for continued use, and if so, continue to use until your replacement arrives. " Ortho will credit and/or replace your remaining inventory of Lot 0048-0964-9287. Please specify full and/or partial sales units remaining and indicate your preference of credit or replacement on your Confirmation of Receipt form. " Complete the enclosed Confirmation of Receipt form no later than June 21, 2019. " Post this notification by each system that processes VITROS GLU Slides or with your user documentation until you switch to an alternate lot. " Please forward this notification if the product was distributed outside of your facility. If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
Quantity in commerce
1099 (5 pack x 60)=329700 slides
Distribution
US Nationwide Distribution - AL, AR, CA, CO, CT, FL, KS, KY, MA, MN, TN
Product codes and lots
Lot Number:0048-0964-9287; Exp Date:2020-11-01 UDI: 10758750009572

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2202-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2202-2019Class IITerminatedVoluntary: Firm initiated