Alinity ci -series System ControlModule, Part Number 03R70 01
Abbott Gmbh & Co. KG began a recall of Alinity ci -series System ControlModule, Part Number 03R70 01 Product Usage: The Alinity ci series System Control Module has a scalable design to provide full... on . FDA classified it as Class II and gives the reason as "Abbott has identified an issue with all on market versions (v2.6.0 and v2.6.1) of Alinity ci series Software where incorrect results may occur after a system Stop due to the Alinity i re use of reaction vessels (RVs)...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2363-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Alinity ci -series System ControlModule, Part Number 03R70 01 Product Usage: The Alinity ci series System Control Module has a scalable design to provide full integration of clinical chemistry and immunoassay analysis. The Alinity ci series System Control Module includes a clinical chemistry module and an immunoassay module, each performing all sample processing activities, and a system control module to provide a single user friendly interface. Each system can be customized by combining one or more sample processing modules, in multiple configurations, with a system control module to form a single workstation. The Alinity i Processing Module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Abbott has identified an issue with all on market versions (v2.6.0 and v2.6.1) of Alinity ci series Software where incorrect results may occur after a system Stop due to the Alinity i re use of reaction vessels (RVs). This issue only occurs if the system is transitioned from Processing to Stopped to Idle.
- Root cause tag
- software
- FDA root cause
- Software in the Use Environment
- Action
- Abbott sent an Urgent Product Correction Notice by letter dated 06/10/2019. The letter identified the affected product, problem and actions to be taken. Abbott has identified an issue with all onmarket versions of Alinity ciseries Software where reuse of reaction vessels (RVs) may occur after a system Stop. This issue only occurs if the system is transitioned from Processing to Stopped to Idle and has the potential to cause incorrect results.The letter provided a mitigation strategy for the consignee to use until the updated software is available. For questions U.S. Customers contact Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service.
- Quantity in commerce
- 1763 units
- Distribution
- Worldwide Distribution - US Nationwide
- Product codes and lots
- All serial numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2363-2019 | Class II | Terminated | Voluntary: Firm initiated |