IDEAL IMPLANT Structured Breast Implant, saline breast implant, in the following sizes: 210cc, 240cc, 270cc...
Ideal Implant Incorporated began a recall of IDEAL IMPLANT Structured Breast Implant, saline breast implant, in the following sizes: 210cc, 240cc, 270cc, 300cc, 335cc, 370cc, 405cc, 440cc, 475cc, 515cc... on . FDA classified it as Class II and gives the reason as "The implants were deflating due to a piece of silicone disrupting the seal of the posterior valve shortly after implantation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2219-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ideal Implant Incorporated
- Product
- IDEAL IMPLANT Structured Breast Implant, saline breast implant, in the following sizes: 210cc, 240cc, 270cc, 300cc, 335cc, 370cc, 405cc, 440cc, 475cc, 515cc, 555cc, 595cc, 635cc, and 675cc.
- Product code
- FWM Prosthesis, Breast, Inflatable, Internal, Saline
- Reason for recall
- The implants were deflating due to a piece of silicone disrupting the seal of the posterior valve shortly after implantation.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The recall was initiated via telephone and email on 3/18/2019 to one consignee and via email on 3/22/2019 to the other consignees and requested return of the implants.
- Quantity in commerce
- 1,704 implants
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NC, ND, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. Country of Canada.
- Product codes and lots
- Serial numbers between 0100292 and 0102777.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2219-2019 | Class II | Terminated | Voluntary: Firm initiated |