ELLIPSE VR, Tiered-therapy cardioverter/defibrillator, REF: CD1411-36Q, UDI: 05414734507738
St Jude Medical Inc. began a recall of ELLIPSE VR, Tiered-therapy cardioverter/defibrillator, REF: CD1411-36Q, UDI: 05414734507738 on . FDA classified it as Class I and gives the reason as "Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2072-2019
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ELLIPSE VR, Tiered-therapy cardioverter/defibrillator, REF: CD1411-36Q, UDI: 05414734507738
- Product code
- LWS Implantable Cardioverter Defibrillator (Non-Crt)
- Reason for recall
- Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 06/20/19, Sales Representatives visited impacted customer accounts to provide talking points that included the reason for recall and to retrieve affected devices. On 06/21/19, hand-delivery of Urgent Medical Device Recall Notices, to physicians supporting implanted patients, commenced. Customers were informed that there is no available option to verify the vulnerability status for implanted devices. Device explant and replacement are recommended. Customers were advised to: 1) Review the device model and serial numbers in the appendix of this letter to identify the impacted patients and return the acknowledgement form to your sales representative; and 2) Device explant and replacement are recommended. The recalling firm will work with customers to provide a replacement device. A copy of this letter is available on https://www.cardiovascular.abbott/us/en/hcp/resources/product/advisories.html. Customers with additional questions were encouraged to call 1-800-727-7846 (Opt3), 8:30am - 5:30pm Central Time, Monday thru Friday.
- Quantity in commerce
- 52
- Distribution
- Worldwide distribution. US nationwide, and countries: DE, FR, IT, ES, NL, PT, SE, HU, DK, PL, IL, GP, DZ, CH, SA, SK, LU, NZ, VN, and NO
- Product codes and lots
- Serial Number: 9828978, 9829577, 9830723, 9830869, 9831468, 9831616, 9831748, 9831795, 9831810, 9831919, 9833545, 9833546, 9834026, 9834051, 9834053, 9834055, 9834059, 9834063, 9834071, 9834140, 9834145, 9834167, 9834171, 9834256, 9834258, 9834259, 9834262, 9834265, 9834271, 9834272, 9834274, 9834275, 9834277, 9834285, 9834289, 9834291, 9834296, 9834299, 9834301, 9834302, 9834329, 9834398, 9834403, 9834407, 9834409, 9834418, 9834422, 9834423, 9834428, 9834431, 9834432, 9834553
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2070-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2071-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2072-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2073-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2074-2019 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2075-2019 | Class I | Terminated | Voluntary: Firm initiated |