Z-2204-2019Class II

LSB Style Anchor Bolt. Labeled with the following parts: 13mm anodized anchor bolt body with anchor bolt cap...

AD-Tech Medical Instrument Corp. / initiated 2019-06-18 / Terminated / root cause: manufacturing process

Ad-Tech Medical Instrument Corporation began a recall of LSB Style Anchor Bolt. Labeled with the following parts: 13mm anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap, 21mm... on . FDA classified it as Class II and gives the reason as "On June 12, 2019 Ad-Tech was made aware that on three different occasions between March 1, 2019 and May 31, 2019 Ad-Tech personnel have distributed non-sterile anchor bolt parts to one independent representative and one...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2204-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
AD-Tech Medical Instrument Corp.
Product
LSB Style Anchor Bolt. Labeled with the following parts: 13mm anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap, 21mm anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap, 13mm non-anodized anchor bolt body with anchor bolt cap containing a green gasket and silicone cap, 26mm non-anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap - Product Usage: The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired to minimize concerns about potential cerebrospinal fluid (CSF) leakage and infection of the subdural space while stabilizing the electrode. Anchor Bolts are secured in the skull to provide an access point for Depth Electrodes.
Product code
GZL Electrode, Depth
Reason for recall
On June 12, 2019 Ad-Tech was made aware that on three different occasions between March 1, 2019 and May 31, 2019 Ad-Tech personnel have distributed non-sterile anchor bolt parts to one independent representative and one international distributor, who may have subsequently sterilized the non-sterile anchor bolts for clinical use. Ad-Tech does not have regulatory clearance to distribute non-sterile anchor bolts and does not have a validated hospital sterilization method for these devices. Further, Ad-Tech labeling does not include instructions for sterilizing anchor bolts.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Ad-Tech Medical notified customers on about 06/18/2019 via "URGENT: MEDICAL DEVICE RECALL" letter. Customers were instructed to immediately examine inventory and quarantine any parts subject to the recall, notify all customers if the affected parts have been further distributed, and provide a copy of the recall notification letter. Additionally, customers were instructed to complete and return the customer response form, contact Ad-Tech Customer Support Specialists to obtain a Return Material Authorization (RMA) number, and return all affected parts to Ad-Tech Medical Instrument Corporation at 400 West Oakview Parkway, Oak Creek, WI 53154. The customer response form should be sent to the Ad-Tech Regulatory Team at FAX: 262-634-5668, Telephone: 262-634-1555, Email: [email protected]. Questions and concerns can also be directed to the Ad-Tech Regulatory Team at the above contact information Monday through Friday, 7:00 AM to 5:00 PM, Central Time.
Quantity in commerce
201 devices
Distribution
Worldwide distribution - US Nationwide distribution in the state of Georgia and country of Australia.
Product codes and lots
All codes shipped between March 1, 2019 and May 31, 2019.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2204-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2204-2019Class IITerminatedVoluntary: Firm initiated