Z-1964-2019Class II
Refurbish RIC5-9-D Probe ROHS compliant Ultrasound Probe RIC5-9-D
GE Healthcare / initiated 2019-07-02 / Terminated / root cause: design
GE Healthcare, LLC began a recall of Refurbish RIC5-9-D Probe ROHS compliant Ultrasound Probe RIC5-9-D on . FDA classified it as Class II and gives the reason as "Potential for a transducer mis-alignment in certain transvaginal probes". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1964-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Refurbish RIC5-9-D Probe ROHS compliant Ultrasound Probe RIC5-9-D
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Potential for a transducer mis-alignment in certain transvaginal probes.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- GE Healthcare notified customers on about 07/02/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that they could continue to use the affected probes, but prior to performing a biopsy, customers should make sure they follow the instructions for safe use in the Basic User Manual to ensure proper device alignment, specifically including the biopsy, biopsy safety and biopsy setup sections. GE Healthcare will correct all affected devices at no cost and a GE Healthcare representative will contact customers to arrange for the correction. Customers were also instructed to complete and return the Medical Device Notification Acknowledgement Response form. Questions or concerns regarding the notification can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity in commerce
- 3 devices
- Distribution
- Distribution in the US to Texas. International distribution to France, Greece and Switzerland.
- Product codes and lots
- Model KTZ303987-R - Serial Number (UDI Number): 226914KR0 (01008406821084301118032721226914KR0), 198026KR7 (Not Available), 220616KR7 (01008406821084301117091521220616KR7), 205375KR9 (01008406821085531116071121205375KR9)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1964-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1965-2019 | Class II | Terminated | Voluntary: Firm initiated |