Z-1965-2019Class II

Repair RIC5-9A-RS System, imaging, pulsed doppler, ultrasonic and Refurbish RIC-5-9-D REALTIME 4D

GE Healthcare / initiated 2019-07-02 / Terminated / root cause: design

GE Healthcare, LLC began a recall of Repair RIC5-9A-RS System, imaging, pulsed doppler, ultrasonic and Refurbish RIC-5-9-D REALTIME 4D on . FDA classified it as Class II and gives the reason as "Potential for a transducer mis-alignment in certain transvaginal probes". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1965-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Repair RIC5-9A-RS System, imaging, pulsed doppler, ultrasonic and Refurbish RIC-5-9-D REALTIME 4D
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
Potential for a transducer mis-alignment in certain transvaginal probes.
Root cause tag
design
FDA root cause
Device Design
Action
GE Healthcare notified customers on about 07/02/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that they could continue to use the affected probes, but prior to performing a biopsy, customers should make sure they follow the instructions for safe use in the Basic User Manual to ensure proper device alignment, specifically including the biopsy, biopsy safety and biopsy setup sections. GE Healthcare will correct all affected devices at no cost and a GE Healthcare representative will contact customers to arrange for the correction. Customers were also instructed to complete and return the Medical Device Notification Acknowledgement Response form. Questions or concerns regarding the notification can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in commerce
2 devices
Distribution
Distribution in the US to Texas. International distribution to France, Greece and Switzerland.
Product codes and lots
Model H48701EJ; Serial Number 205375KR9; (UDI Number) 01008406821085531116071121205375KR9.   Model KTZ157043-R Serial Number 152323KR2 (UDI Number Not Available)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1965-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1964-2019Class IITerminatedVoluntary: Firm initiated
Z-1965-2019Class IITerminatedVoluntary: Firm initiated