Z-2091-2019Class II
Monaco RTP System, software versions 5.10 and 5.11, Radiation Treatment Planning System
Elekta Limited / initiated 2019-07-03 / Open, Classified
Elekta Inc began a recall of Monaco RTP System, software versions 5.10 and 5.11, Radiation Treatment Planning System on . FDA classified it as Class II and gives the reason as "When creating 3D plans using either MU or Dose weighting modes, if the user changes the number of fractions, rescales the plan and then changes the wedge angle, the Monitor Units are scaled incorrectly". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2091-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Elekta Limited
- Product
- Monaco RTP System, software versions 5.10 and 5.11, Radiation Treatment Planning System
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- When creating 3D plans using either MU or Dose weighting modes, if the user changes the number of fractions, rescales the plan and then changes the wedge angle, the Monitor Units are scaled incorrectly.
- FDA root cause
- Under Investigation by firm
- Action
- Elekta began distribution of their Important Field Safety Notice to all affected customers from 07/03/2019 by email. The notice informs users of the specific product and version numbers affected by the issue, and any work-arounds that can be used to avoid the issue. The problem will be resolved in a future Monaco software release.
- Quantity in commerce
- 1396 units
- Distribution
- worldwide
- Product codes and lots
- all Monaco Radiation Treatment Planning units installed with software version 5.10 or 5.11
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2091-2019 | Class II | Ongoing | Voluntary: Firm initiated |