IMRIS MR/X-ray head fixation device, Model HFD200
Deerfield Imaging, Inc. began a recall of IMRIS MR/X-ray head fixation device, Model HFD200 on . FDA classified it as Class II and gives the reason as "Breaking of the HFD200 Drive Knob force indicator due to over-torqueing of the drive knob beyond the maximum allowable pinning force specification may result in the clinical staff unknowingly pinning the patient's skull...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0163-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- IMRIS Imaging, Inc.
- Product
- IMRIS MR/X-ray head fixation device, Model HFD200
- Product code
- HBL Holder, Head, Neurosurgical (Skull Clamp)
- Reason for recall
- Breaking of the HFD200 Drive Knob force indicator due to over-torqueing of the drive knob beyond the maximum allowable pinning force specification may result in the clinical staff unknowingly pinning the patient's skull with insufficient force; the clinical staff unknowingly pinning the patient's skull with excessive force; or the clinical staff identifying the issue and not being able to continue with the procedure.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm disseminated a field safety notice on 06/26/2019 by letter. The notice explained the problem and requested that the consignee inspect the skull clamp prior to use and be aware that overtightening of the drive knob may result in failure. The drive knobs will be replaced with an updated design when available.
- Quantity in commerce
- 3 units
- Distribution
- TN, NH, FL
- Product codes and lots
- HFD Serial Number Drive Knob Serial Number 10002481 10003305 10002351 10005723 10004247 10002448
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0163-2020 | Class II | Terminated | Voluntary: Firm initiated |