SenTec Membrane Changer Set (1 charger plus 1 insert) -
SenTec AG began a recall of SenTec Membrane Changer Set (1 charger plus 1 insert) - Product Usage: is indicated for continuous, non-invasive monitoring of carbon dioxide tension and... on . FDA classified it as Class II and gives the reason as "There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2155-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sentec AG
- Product
- SenTec Membrane Changer Set (1 charger plus 1 insert) - Product Usage: is indicated for continuous, non-invasive monitoring of carbon dioxide tension and oxygen tension as well as oxygen saturation and pulse rate in adult and pediatric patients. In neonatal patients the SDMS is indicated for carbon dioxide and oxygen tension monitoring only. Oxygen tension monitoring is contraindicated for patients under gas anesthesia.
- Product code
- LKD Monitor, Carbon-Dioxide, Cutaneous
- Reason for recall
- There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On May 2, 2019, the firm sent Guidance for SenTec Distributors to their distributors advising them of the issue, requesting them to check their stock and notify their end users via a voluntarily initiated field correction action associated with Membrane Changers and their Inserts which was enclosed with the "recall" letters. The distributors were to report back to SenTec via email.
- Quantity in commerce
- 3 sets
- Distribution
- Worldwide distribution - US Nationwide in the state of Missouri and countries of Argentina, Australia, Austria, Belgium, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, Ecuador, Europe, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Indonesia, Israel, Italy, Kuwait, Malaysia, Finland, Portugal, Romania, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, United Arab Emirates and United Kingdom.
- Product codes and lots
- Lots: 180961, 181097, and 190053
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2152-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2153-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2154-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2155-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2156-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
