Z-2341-2019Class II

N Cuvette Segments for BN ProSpec System-IVD a system for the quantitative determination of proteins in...

Siemens Healthcare Diagnostics Inc. / initiated 2019-06-27 / Terminated

Siemens Healthcare Diagnostics, Inc. began a recall of N Cuvette Segments for BN ProSpec System-IVD a system for the quantitative determination of proteins in serum, plasma, urine and CSF. Siemens Material Number... on . FDA classified it as Class II and gives the reason as "Iincreased (falsely elevated) Albumin results compared to expected results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2341-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
N Cuvette Segments for BN ProSpec System-IVD a system for the quantitative determination of proteins in serum, plasma, urine and CSF. Siemens Material Number (SMN):10446539
Product code
JZW Nephelometer
Reason for recall
Iincreased (falsely elevated) Albumin results compared to expected results
FDA root cause
Process change control
Action
Siemens Healthcare Diagnostics issued Urgent Medical Device Correction (UMDC) letter PP-19-002.A.US to the US customers June 27, 2019 and May 2019 via FedEx overnight delivery and email to regional countries for implementation Outside the US (OUS). Customers were instructed to contact their Siemens Healthineers Customer Care Center for further follow-up in case they experienced sporadically increased Albumin results compared to expected results. Customers were also instructed to complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days.
Quantity in commerce
664 units
Distribution
Nationwide
Product codes and lots
Lot # 5890128326 and Lot # 182670004 UDI: 00842768012334

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2341-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2341-2019Class IITerminatedVoluntary: Firm initiated