Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2234-2019Class II

Orthopedic Salvage System (OSS) Hybrid Poly Tibia Item # 151813

Zimmer GmbH / initiated 2019-07-02 / Terminated / root cause: manufacturing process

Zimmer Biomet, Inc. began a recall of Orthopedic Salvage System (OSS) Hybrid Poly Tibia Item # 151813 on . FDA classified it as Class II and gives the reason as "Cleaning processes potentially being ineffective". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2234-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Orthopedic Salvage System (OSS) Hybrid Poly Tibia Item # 151813
Product code
JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Reason for recall
Cleaning processes potentially being ineffective
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Certificate of Acknowledgement and send to Corporate Quality. [email protected]. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Certificate of Acknowledgement and send to [email protected]. This form must be returned. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed t
Quantity in commerce
39
Distribution
Distributed to the following states in the USA: AZ, CA, DC, GA, IL, KY, LA, MA, MD, MN, NC, NJ, NM, OH,OK, OR, SC, TX. Distributed to Japan and Netherlands.
Product codes and lots
Lot # All lots expiring prior to 06/30/2024 Affected products were distributed between June 2016 and February 2019.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2234-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2231-2019Class IITerminatedVoluntary: Firm initiated
Z-2232-2019Class IITerminatedVoluntary: Firm initiated
Z-2233-2019Class IITerminatedVoluntary: Firm initiated
Z-2234-2019Class IITerminatedVoluntary: Firm initiated
Z-2235-2019Class IITerminatedVoluntary: Firm initiated
Z-2236-2019Class IITerminatedVoluntary: Firm initiated