14F x 30CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061425
Medical Components, Inc dba MedComp began a recall of 14F x 30CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061425 Product Usage: An aid for introducing a hemodialysis catheter on . FDA classified it as Class II and gives the reason as "Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2217-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medical Components, Inc.
- Product
- 14F x 30CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061425 Product Usage: An aid for introducing a hemodialysis catheter
- Product code
- LFJ Catheter, Subclavian
- Reason for recall
- Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- On March 13, 2019, the firm notified customers via Urgent Medical Device Recall letter. Customers were informed that the kits contain a 15F "non-valved" peelable introducer despite the label indicating that the kit is packaged with a "valved" introducer. Customers were asked to do the following: 1. Immediately examine your inventory and quarantine product subject to recall. 2. If you have further distributed this product, please identify your customers and immediately notify them of the recall. 3. Contact your Medical Components, Inc. customer service representative for a Returned Goods Authorization (RGA) number if you have product to return. The phone number to call is 215-256-4201. 4.Complete and return the Medical Device Recall Return Response Form. The form can be returned by fax (215-256-9191) or email ([email protected]). Contact Information for questions: Monday through Friday, 8:00 AM to 4:00 PM, Eastern Standard Time. Phone: 215-256-4201; Email: [email protected]
- Quantity in commerce
- 6830
- Distribution
- US Nationwide Distribution CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
- Product codes and lots
- MCAM240 MCBL480 MCBS870 MCCF480 MCCJ320 MCCW690 MCCZ500 MCFG830 MCFR160 MCFS760 MCFZ360 MKAJ330 MKAR130 MKBL500 MKBX060 MMAB730
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2215-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2216-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2217-2019 | Class II | Terminated | Voluntary: Firm initiated |