Z-2185-2019Class II

VITROS Chemistry Products Calibrator Kit 32, for use with VITROS clinical chemistry analyzer

Ortho-Clinical Diagnostics, Inc. / initiated 2019-07-08 / Terminated / root cause: software

Ortho Clinical Diagnostics began a recall of VITROS Chemistry Products Calibrator Kit 32, for use with VITROS clinical chemistry analyzer on . FDA classified it as Class II and gives the reason as "Potential Unsuccessful Calibration Due to Software Anomaly in VITROS System". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2185-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products Calibrator Kit 32, for use with VITROS clinical chemistry analyzer
Product code
JIX Calibrator, Multi-Analyte Mixture
Reason for recall
Potential Unsuccessful Calibration Due to Software Anomaly in VITROS System
Root cause tag
software
FDA root cause
Software design
Action
Customers were notified prior to July 10, 2019. For the VITROS 250, 250AT, and 350 chemistry system: Do not use VITROS Calibrator Kit 32 to calibrate assays until an Ortho trained service personnel reconfigures your system(s) with the correct parameters for VITROS Calibrator Kit 32. Complete the enclosed Confirmation of Receipt form no later than July 15, 2019. Post this notification by each system that may use VITROS Calibrator Kit 32 until your system is reconfigured by an Ortho trained representative. Please forward this notification if the product was distributed outside of your facility. For the VITROS 4600,5600 and 5,1 FS chemistry System: If your system is configured with User Modified parameters, choose an option provided in the Interim Resolutions section listed above. Install ADD DRV 6040 or above upon availability. Complete the enclosed Confirmation of Receipt form no later than July 15, 2019. Post this notification either by each system that processes VITROS Calibrator Kit 32 or with your user documentation until DRV 6040 or above is installed on your system(s). Please forward this notification if the product was distributed outside of your facility. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Brazil, Canada, Chile, France, India, and Philippines.
Product codes and lots
product code:6844508 UDI: 10758750032105 Affected Assay Data Disk (ADD) DRV No: DRV 6032 or above

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2185-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2174-2019Class IITerminatedVoluntary: Firm initiated
Z-2175-2019Class IITerminatedVoluntary: Firm initiated
Z-2176-2019Class IITerminatedVoluntary: Firm initiated
Z-2177-2019Class IITerminatedVoluntary: Firm initiated
Z-2178-2019Class IITerminatedVoluntary: Firm initiated
Z-2179-2019Class IITerminatedVoluntary: Firm initiated
Z-2180-2019Class IITerminatedVoluntary: Firm initiated
Z-2181-2019Class IITerminatedVoluntary: Firm initiated
Z-2182-2019Class IITerminatedVoluntary: Firm initiated
Z-2183-2019Class IITerminatedVoluntary: Firm initiated
Z-2184-2019Class IITerminatedVoluntary: Firm initiated
Z-2185-2019Class IITerminatedVoluntary: Firm initiated