Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2160-2019Class II

Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog...

Howmedica Osteonics Corp / initiated 2019-06-18 / Terminated / root cause: labeling and UDI

Howmedica Osteonics Corp. began a recall of Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100 on . FDA classified it as Class II and gives the reason as "Discrepant length; Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2160-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
Discrepant length; Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
1. Please note this product has already been removed by your local Stryker Sales Representative and/or Stryker Branch or Agency. This communication is being made for notification purposes only; there are no additional product return requirements for this matter. 2. Please assist us in meeting our regulatory obligation by completing and signing the enclosed Product Recall Acknowledgment Form and email a copy to Stryker Spine at [email protected] or fax the form to 201-575-4675. If you have any questions, please contact Christa Joisil or Matthew Kelleher from the Regulatory Compliance Team: [email protected] or 201.749.8090
Quantity in commerce
11
Distribution
US Nationwide Distribution - CA, FL, GA, MA, NY, PA, TX
Product codes and lots
Lot # B88751 UDI: (01)07613327351576(11)190125(10)B88751

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2160-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2160-2019Class IITerminatedVoluntary: Firm initiated