Berchtold Chromophare F628 Surgical Lights
Stryker Communications began a recall of Berchtold Chromophare F628 Surgical Lights. Product Usage: The medical lights are for usage in hospital treatment rooms. They provide localized lighting... on . FDA classified it as Class II and gives the reason as "The back cover of the light head was insufficiently assembled and could potentially fall off the light". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2540-2019
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- STRYKER Communications, Inc.
- Product
- Berchtold Chromophare F628 Surgical Lights. Product Usage: The medical lights are for usage in hospital treatment rooms. They provide localized lighting focused on the patient s body for detecting and treating disease, injury, and disability. The surgical lights may only be used in rooms used for medical purposes that have been equipped in accordance with VDE 0100-710 or IEC 60364-7-10 in accordance with the regulations.
- Product code
- FSY Light, Surgical, Ceiling Mounted
- Reason for recall
- The back cover of the light head was insufficiently assembled and could potentially fall off the light.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The recalling firm issued letters via FedEx on 3/22/2019 to their U.S. consignees. The recall letter identified the affected product, problem and actions to be taken. ,The letter instructs the customer to disseminate it within their organization and that a Stryker representative will begin replacements for customer facilities in January. For the replacement, the Stryker representative will perform an onsite rework of the light head cover. In the meantime, the lights can continue to remain in full use. A response form was enclosed for the customer to indicate they have read and understood the instructions. The form was to be returned via email. The recalling firm's representatives will contact the customer to schedule an inspection of their light and will track completion of the visits and inspections.
- Quantity in commerce
- 3,344 lights
- Distribution
- Distribution was nationwide, including to Puerto Rico. There was also government/military distribution. Foreign distribution was made to Argentina, Austria, Australia, Belgium, Bulgaria, Canada, China, Colombia, Denmark, France, Germany, Hong Kong, India, Indonesia, Italy, Japan, Kazakhstan, Korea, Kuwait, Lithuania, Mexico, Netherlands, Philippines, Poland, Russia, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.
- Product codes and lots
- Serial number ranges: 83171-0017818 to 83171-0123938
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2539-2019 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2540-2019 | Class II | Ongoing | Voluntary: Firm initiated |