Azurion lnterventional Fluoroscopic X-ray System, with software version 1.2
Philips Medical Systems Nederlands began a recall of Azurion lnterventional Fluoroscopic X-ray System, with software version 1.2 on . FDA classified it as Class II and gives the reason as "Philips has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with 21 CFR 1003.2(b)(1). When performing Cine runs, the Azurion's automatic exposure...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1997-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Product
- Azurion lnterventional Fluoroscopic X-ray System, with software version 1.2
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Philips has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with 21 CFR 1003.2(b)(1). When performing Cine runs, the Azurion's automatic exposure control software may set technique factors that result in radiation emissions too low to obtain useful diagnostic images.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On November 30, 2018, Philips initiated a software update in the field. Among other performance improvements , this software update included the correction for poor image quality under very specific situations.
- Quantity in commerce
- 229
- Distribution
- Nationwide
- Product codes and lots
- software version R1.2
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1997-2019 | Class II | Terminated | FDA Mandated |