Z-1997-2019Class II

Azurion lnterventional Fluoroscopic X-ray System, with software version 1.2

Philips Medical Systems Nederland B.V. / initiated 2018-11-30 / Terminated / root cause: design

Philips Medical Systems Nederlands began a recall of Azurion lnterventional Fluoroscopic X-ray System, with software version 1.2 on . FDA classified it as Class II and gives the reason as "Philips has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with 21 CFR 1003.2(b)(1). When performing Cine runs, the Azurion's automatic exposure...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1997-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips Medical Systems Nederland B.V.
Product
Azurion lnterventional Fluoroscopic X-ray System, with software version 1.2
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Philips has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with 21 CFR 1003.2(b)(1). When performing Cine runs, the Azurion's automatic exposure control software may set technique factors that result in radiation emissions too low to obtain useful diagnostic images.
Root cause tag
design
FDA root cause
Device Design
Action
On November 30, 2018, Philips initiated a software update in the field. Among other performance improvements , this software update included the correction for poor image quality under very specific situations.
Quantity in commerce
229
Distribution
Nationwide
Product codes and lots
software version R1.2

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1997-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1997-2019Class IITerminatedFDA Mandated