This is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays...
Randox Laboratories Ltd. began a recall of This is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays. Randox calibration sera contain 42 analytes and... on . FDA classified it as Class III and gives the reason as "Randox Laboratories realigned RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target for the Randox Calibration Serum Level 3 (for calibrator lot 961UE...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2490-2019
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Randox Laboratories, Limited
- Product
- This is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays. Randox calibration sera contain 42 analytes and are based on lyophilized human serum. The concentrations and activities are suitable for calibration of clinical chemistry assays both manually and on a wide range of automatic analyzers. Constituent concentrations are available at 2 levels.
- Product code
- JIX Calibrator, Multi-Analyte Mixture
- Reason for recall
- Randox Laboratories realigned RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target for the Randox Calibration Serum Level 3 (for calibrator lot 961UE onwards).
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- Randox Laboratories Ltd. issued Urgent Medical Device Correction letter on 7/12/19 advising users of the problem, health risk and action to take: Replace all Quality Control value sheets with the revised sheets available on www.randox.com . Discuss the contents of this notice with your Medical Director. Complete and return the response form 12187-QA to [email protected] within five working days.
- Quantity in commerce
- 307 kits
- Distribution
- Nationwide Foreign: Antigua, Aruba, Bahamas, Granada, Tobago,
- Product codes and lots
- Lot Numbers: 961UE onwards GTN: 05055273200966
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2490-2019 | Class III | Terminated | Voluntary: Firm initiated |