Z-2490-2019Class III

This is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays...

Randox Laboratories, Limited / initiated 2019-07-12 / Terminated / root cause: manufacturing process

Randox Laboratories Ltd. began a recall of This is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays. Randox calibration sera contain 42 analytes and... on . FDA classified it as Class III and gives the reason as "Randox Laboratories realigned RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target for the Randox Calibration Serum Level 3 (for calibrator lot 961UE...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2490-2019
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Randox Laboratories, Limited
Product
This is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays. Randox calibration sera contain 42 analytes and are based on lyophilized human serum. The concentrations and activities are suitable for calibration of clinical chemistry assays both manually and on a wide range of automatic analyzers. Constituent concentrations are available at 2 levels.
Product code
JIX Calibrator, Multi-Analyte Mixture
Reason for recall
Randox Laboratories realigned RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target for the Randox Calibration Serum Level 3 (for calibrator lot 961UE onwards).
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Randox Laboratories Ltd. issued Urgent Medical Device Correction letter on 7/12/19 advising users of the problem, health risk and action to take: Replace all Quality Control value sheets with the revised sheets available on www.randox.com . Discuss the contents of this notice with your Medical Director. Complete and return the response form 12187-QA to [email protected] within five working days.
Quantity in commerce
307 kits
Distribution
Nationwide Foreign: Antigua, Aruba, Bahamas, Granada, Tobago,
Product codes and lots
Lot Numbers: 961UE onwards GTN: 05055273200966