Z-0195-2020Class II
DEMO No Needle Corrugated Carton 20 U/D, Part Number 2260-03
Valeritas, Incorporated / initiated 2019-07-03 / Terminated / root cause: manufacturing process
Valeritas, Incorporated began a recall of DEMO No Needle Corrugated Carton 20 U/D, Part Number 2260-03 on . FDA classified it as Class II and gives the reason as "There is a potential for "No Needle Demo Units" to contain a needle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0195-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Valeritas, Incorporated
- Product
- DEMO No Needle Corrugated Carton 20 U/D, Part Number 2260-03
- Product code
- LZG Pump, Infusion, Insulin
- Reason for recall
- There is a potential for "No Needle Demo Units" to contain a needle.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- On July 3, 2019, Valeritas issued notices to customers via first class U.S. postal service. The removal strategy is to notify active HCP office accounts from the manufacturing start date of effected product (January 2018) who may have No Needle Demo units in their possession. The removal letter instructs HCP offices to dispose of any No Needle Demo product on hand and acknowledgement by signing and returning the removal letter as instructed.
- Quantity in commerce
- 49,074 devices
- Distribution
- Nationwide within U.S.
- Product codes and lots
- UDI: (01)10385609020039(17)XXXXXX(10)XXXXXXXX Lot Numbers: DM218001 DM218002 DM218003 DM218004 DM218005 DM218006 DM218020 DM218025 DM218034 DM218035 DM218036
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0195-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0196-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0197-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0198-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0199-2020 | Class II | Terminated | Voluntary: Firm initiated |