Z-2336-2019Class II
Kit BD Max Vaginal Panel; Catalog # 443712
Becton Dickinson & Co / initiated 2019-07-17 / Terminated / root cause: manufacturing process
Becton Dickinson & Co. began a recall of Kit BD Max Vaginal Panel; Catalog # 443712 on . FDA classified it as Class II and gives the reason as "The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2336-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- Kit BD Max Vaginal Panel; Catalog # 443712
- Product code
- PQA Vaginitis And Bacterial Vaginosis Nucleic Acid Detection System
- Reason for recall
- The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- 1. Thoroughly inspect all foil bags prior to use. If any of the foil bags in your inventory contain holes, dispose of the product and note the quantity in the attached Customer Response Form in order to request replacement. 2. Even if you do not have inventory of the product listed, complete the attached Customer Response Form and return to the BD contact noted on the form so that BD may acknowledge your receipt of this notification. 3. Share this Advisory Letter with all users of the BD MAX instrument within your facility to ensure awareness. 4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDAs MedWatch Adverse Event Reporting program. Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787 BD Customer/Technical Support, 800-638-8663 Monday Friday between 7:00am and 7:00pm (EST) in the United States. For customers outside the US, contact your local BD representative or distributor.
- Quantity in commerce
- 666 units
- Distribution
- AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, NH, NJ, NM, NV, NY, OH, OR, PA, PR, RI, SC, TN, TX, VA, WA, WI, WV Worldwide: Australia, Belgium, Brazil, Canada, Chile, China, Hong Kong, Japan, Korea, Mexico, Singapore, Taiwan
- Product codes and lots
- Lot # 9092737; UDI (GTIN, DI+PI): (01)00382904437121 (17)200321(10)9092737(30)1; Exp : 3/21/2020 Lot # 9092738; UDI (GTIN, DI+PI): (01)00382904437121 (17)200321(10)9092738(30)1; Exp :3/21/2020 Lot # 9127920; UDI (GTIN, DI+PI): (01)00382904437121 (17)200321(10)9127920(30)1; Exp : 3/21/2020
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2321-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2322-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2323-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2324-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2325-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2326-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2327-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2328-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2329-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2330-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2331-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2332-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2333-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2334-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2335-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2336-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2337-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2338-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2339-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.