Z-1591-2020Class II

Medtronic MiniMed Paradigm Insulin pump, model number MMT-523

Medtronic, Inc. / initiated 2019-06-27 / Open, Classified / root cause: design

Medtronic MiniMed began a recall of Medtronic MiniMed Paradigm Insulin pump, model number MMT-523 on . FDA classified it as Class II and gives the reason as "The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1591-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Medtronic MiniMed Paradigm Insulin pump, model number MMT-523
Product code
OYC Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor
Reason for recall
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.
Root cause tag
design
FDA root cause
Device Design
Action
On June 27, 2019, Medtronic issued an Urgent Field Safety Notification letter via mail and email to its consignees. Medtronic made the decision to renotify customers, and on January 4, 2023, Medtronic issued an Urgent Medical Device Correction - Update letter. The communication asks consignees to take the following actions: .To find the software version for the MiniMed Paradigm pumps, go to the STATUS screen: . To open the STATUS screen, press ESC until the STATUS screen appears. . To view more text on the STATUS screen, press the up or down arrow to scroll and view all the information. . To exit the STATUS screen, press ESC until the STATUS screen disappears. ACTIONS REQUIRED: 1. Review and take the following cybersecurity precautions listed below. 2. Please complete and sign the customer confirmation form attached as indicated and return by mail using the attached prepaid envelope. You can also email this form to our Field Corrective Action Department at [email protected] or send via fax to 1-888-844-8084 to the attention of FCA Department.
Quantity in commerce
23,221
Distribution
US nationwide distribution including Puerto Rico.
Product codes and lots
All software versions 2.4A or lower

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1591-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1581-2020Class IIOngoingVoluntary: Firm initiated
Z-1582-2020Class IIOngoingVoluntary: Firm initiated
Z-1583-2020Class IIOngoingVoluntary: Firm initiated
Z-1584-2020Class IIOngoingVoluntary: Firm initiated
Z-1585-2020Class IIOngoingVoluntary: Firm initiated
Z-1586-2020Class IIOngoingVoluntary: Firm initiated
Z-1587-2020Class IIOngoingVoluntary: Firm initiated
Z-1588-2020Class IIOngoingVoluntary: Firm initiated
Z-1589-2020Class IIOngoingVoluntary: Firm initiated
Z-1590-2020Class IIOngoingVoluntary: Firm initiated
Z-1591-2020Class IIOngoingVoluntary: Firm initiated
Z-1592-2020Class IIOngoingVoluntary: Firm initiated
Z-1593-2020Class IIOngoingVoluntary: Firm initiated
Z-1594-2020Class IIOngoingVoluntary: Firm initiated
Z-1595-2020Class IIOngoingVoluntary: Firm initiated
Z-1596-2020Class IIOngoingVoluntary: Firm initiated