Arrow AGB+ Multi-Lumen CVC Kit-catheter permits venous access to central circulation by way of subclavian...
Arrow International Inc began a recall of Arrow AGB+ Multi-Lumen CVC Kit-catheter permits venous access to central circulation by way of subclavian, jugular, and femoral veins Product Code... on . FDA classified it as Class II and gives the reason as "Arrow AGB+ Multi-Lumen CVC Kit - lidstock incorrectly states within CDC-45703-1A a catheter 20cm in length. The banner card correctly states the product is CDC-42703-1A with a catheter 16cm in length". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0192-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- Arrow AGB+ Multi-Lumen CVC Kit-catheter permits venous access to central circulation by way of subclavian, jugular, and femoral veins Product Code: CDC-42703-1A - Product Usage: The ARROWg+ard Blue PLUS catheter permits venous access to central circulation by way of subclavian, jugular, and femoral veins. The ARROWg+ard technology is intended to provide protection against catheter-related bloodstream infections (CRBSIs). It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use (> 30 days). Clinical effectiveness of the ARROWg+ard Blue PLUS catheter in preventing CRBSIs compared to the original ARROWg+ard Blue catheter has not been studied.
- Product code
- FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
- Reason for recall
- Arrow AGB+ Multi-Lumen CVC Kit - lidstock incorrectly states within CDC-45703-1A a catheter 20cm in length. The banner card correctly states the product is CDC-42703-1A with a catheter 16cm in length
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Teleflex issued notification letter dated July 17, 2019 issued via FedEx 2-day mail identifies the problem, health risk and action to take: 1. If you have affected stock in inventory, immediately discontinue use and quarantine the products. 2. Please complete the enclosed Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to [email protected]. If you do have stock a customer service representative will contact you to arrange issuing correct labels to relabel the product. Questions, feel free to contact your local sales representative or Customer Service at 1-866-396-2111.
- Quantity in commerce
- 745 eaches
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Lot Number: 13F18L0759
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0192-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY