Z-0042-2020Class II
Online TDM Vancomycin Gen.3 (100 tests) Catalog # 06779336190 Analyzer/Module: cobas c 311 and cobas c 501/502
Roche Diagnostics GmbH / initiated 2019-07-23 / Terminated
Roche Diagnostics Operations, Inc. began a recall of Online TDM Vancomycin Gen.3 (100 tests) Catalog # 06779336190 Analyzer/Module: cobas c 311 and cobas c 501/502 on . FDA classified it as Class II and gives the reason as "Inaccurately Low Sample Results". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0042-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Online TDM Vancomycin Gen.3 (100 tests) Catalog # 06779336190 Analyzer/Module: cobas c 311 and cobas c 501/502
- Product code
- LEH Radioimmunoassay, Vancomycin
- Reason for recall
- Inaccurately Low Sample Results
- FDA root cause
- Under Investigation by firm
- Action
- 1. Consignees will be provided instruction to download the updated e-barcode either from the cobas link or from the cobas e-library when available. 2. If analyzers cobas link is not connected to the Roche network, a future version of the cobas e-library CD will contain the revised e-barcodes. Download the e-barcode when available. 3. If analyzers cobas link is not connected to the Roche network and they require the updated VANC3 application e-barcode before the cobas e-library CD is available, they will be instructed to contact the Roche Support Network Customer Support Center at 1-800-428-2336 to request a Field Application Specialist to provide the VANC3 updated application. 4. Calibrate and run quality controls (QC) before running samples. 5.Complete all sections of the enclosed fax form. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
- Quantity in commerce
- 27676 units
- Distribution
- AK AL AR AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY
- Product codes and lots
- All Lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0042-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0043-2020 | Class II | Terminated | Voluntary: Firm initiated |