Z-0171-2020Class II

Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump

Abbott Medical / initiated 2018-08-01 / Terminated / root cause: software

ICU Medical Inc began a recall of Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump on . FDA classified it as Class II and gives the reason as "A software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL). When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0171-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump.
Product code
FRN Pump, Infusion
Reason for recall
A software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL). When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User Interface.
Root cause tag
software
FDA root cause
Software change control
Action
On 8/1/2018 a "Urgent Medical Device Correction" letter was mailed to consignees via traceable mailing. The medical device correction letter informed consignees of the following: 1. To reduce the likelihood of the issue occurring at the bedside, you can consider not utilizing the option to "Force Pump to Accept Drug Library During Power Down" when finalizing the CDL with MedNet. In doing so, when powering down the infuser the clinician will have the option to install the CDL or not, based on the clinical situation. 2. ICU Medical will release a software update to correct the issue and will contact you when the update is available. 3. . Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular 4.Please inform potential users in your organization of this notification, complete the attached response form and return it to the fax number (1-888-628-6849 or e-mail address ([email protected]) on the form, even if you do not have the affected product. 5. For further inquiries, please contact ICU Medical using the information provided below: Global Complaint Management - 1-844-654-7780 or [email protected] Technical Support Center - 1-800-241-4002, option 4 or [email protected]
Quantity in commerce
47,152 devices
Distribution
US nationwide distribution, and Canada
Product codes and lots
All Plum 360 Infusers with software versions 15.10.00.010/.011/.012 and 15.11.00.017 when used with an ICU Medical MedNet CDL

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0171-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0171-2020Class IITerminatedVoluntary: Firm initiated