Z-1906-2020Class II
da Vinci SP Instrument Arm Drape. REF 430013-12 QTY 1 Pouch
Intuitive Surgical, Inc. / initiated 2018-10-25 / Terminated / root cause: design
Intuitive Surgical, Inc. began a recall of da Vinci SP Instrument Arm Drape. REF 430013-12 QTY 1 Pouch on . FDA classified it as Class II and gives the reason as "The instrument drape may tear while draping the Patient Side Cart presenting a breach in the sterile barrier". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1906-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- da Vinci SP Instrument Arm Drape. REF 430013-12 QTY 1 Pouch
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- The instrument drape may tear while draping the Patient Side Cart presenting a breach in the sterile barrier.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 10/25/2018, the firm sent an "Urgent Medical Device Correction" Notification Letter to Customers via FedEx (to OUS customers - emailed and hand-delivered) are asked to take actions during draping to avoid snagging the drape during installation and to continue following the instructions in the System User Manual. Representatives will be available by phone to answer questions related to the device correction. Acknowledgement form included with the recall letter is requested to be completed and returned to ISI asking that appropriate personnel be informed about the device correction.
- Quantity in commerce
- 1818 pouches
- Distribution
- Updated information provided on 03/18/2020: FL, OH, TN, NJ, PA, MD, NY, CA, MN, NV, OR, HI, VA, TX, AL, CA, IL, CN, and NC. OUS: APAC (Asia Pacific) - South Korea
- Product codes and lots
- Model Number: 430013-12 UDI Number: 00886874113486 Lot Numbers: All Lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1906-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.