Z-2447-2019Class II

Flexible Intubation Fiberscope, Model # 11301AB1

Karl Storz Endoscopy-America, Inc. / initiated 2019-04-24 / Terminated / root cause: manufacturing process

Karl Storz Endoscopy began a recall of Flexible Intubation Fiberscope, Model # 11301AB1 on . FDA classified it as Class II and gives the reason as "The action is being initiated due to intubation scopes improperly manufactured with a double port luer instead of a single port luer". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2447-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Karl Storz Endoscopy-America, Inc.
Product
Flexible Intubation Fiberscope, Model # 11301AB1
Product code
EOQ Bronchoscope (Flexible Or Rigid)
Reason for recall
The action is being initiated due to intubation scopes improperly manufactured with a double port luer instead of a single port luer.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm mailed a "safety alert" on April 24, 2019 via FedEx. 2nd day delivery. The safety alert advised customers to do the following: 1. If you use the following automated processes, the high level disinfection/liquid chemical sterilization of the flexible endoscope cannot be guaranteed: Reprocessing Method Impact, Medivators AER * Patient Safety may be Compromised, STERIS System 1E (SS1E)** Patient Safety may be Compromised The above connectors can only connect onto one port, therefore leaving the other port open. Depending on the reprocessing system, the cycle could result in an error or in successful completion of the cycle. " If an error code is displayed, the user would know the cycle is not complete and the scope has not been successfully reprocessed. " If the cycle is completed without an error code, the high-level disinfection/liquid chemical sterilization of the flexible endoscope cannot be guaranteed. Improper reprocessing may result in incomplete decontamination, which may result in a cross-contamination patient infection. If you use Medivators AER or STERIS SS1E to reprocess this scope, please be aware of the possible risks and take the necessary actions to address any patient safety concerns. 2. Please complete the attached Action Requested form and follow the instructions outlined on the form. 3. you have any questions please contact the following person: This Recall Action, Emily Cuadros, Complaint Handling , 424-218-8289 Reprocessing, Shaun McGinley, Director of Reprocessing, 424-218-8247 Patient Safety, Dr. Dennis Fowler, Chief Medical Officer, 424-218-8603
Quantity in commerce
14 scopes
Distribution
US: VA, MO, CA, OH, NM, NY, CO, LA, ME, OUS: Canada
Product codes and lots
Lot numbers: 2183827, 2195794, 2215741, 2229258, 2235373, 2248169, 2193769

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2447-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2447-2019Class IITerminatedVoluntary: Firm initiated