Z-2440-2019Class II

syngo.CT software version VA48A_SP5.5 in the following SOMATOM computed tomography models: SOMATOM Definition...

Siemens Medical Solutions USA, Inc. / initiated 2019-07-25 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of syngo.CT software version VA48A_SP5.5 in the following SOMATOM computed tomography models: SOMATOM Definition Edge-Model #10590000; SOMATOM Definition AS Model... on . FDA classified it as Class II and gives the reason as "SOMATOM Definition Edge, SOMATOM Definition AS, SOMATOM Definition Flash with software syngo.CT with Software version VA48A_SP5 may result in scanning workflow interruptions and unexpected user notifications. and may...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2440-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
syngo.CT software version VA48A_SP5.5 in the following SOMATOM computed tomography models: SOMATOM Definition Edge-Model #10590000; SOMATOM Definition AS Model #8098027; SOMATOM Definition Flash Model #10430603
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
SOMATOM Definition Edge, SOMATOM Definition AS, SOMATOM Definition Flash with software syngo.CT with Software version VA48A_SP5 may result in scanning workflow interruptions and unexpected user notifications. and may result in a delay in diagnosis and/or patient rescans
Root cause tag
software
FDA root cause
Software design
Action
Siemens issued notification Urgent Medical Device Correction letter on 7/25/19 issued via electronic mail to consignees or by USPS certified mail if a consignee email address was not available. Letter states problem, health risk, action to take: The remote software update installation process will require approximately 30 minutes for completion. The process is completely automatic; make sure the system and power are stable before and during the process and do not switch off the system during the update process. Siemens highly recommends starting the installation when the scanner will not be in use or when the necessary time for the update to be completed can be scheduled. If your system is connected to Siemens Remote Connection, this update will be performed remotely. Following the corrective action, the cause will be eliminated and a recurrence of the identified issues is prevented. If you have any questions, please contact our service organization at 1-800-888-7436.
Quantity in commerce
2867 units worldwide (OUS); 1464 units US
Distribution
Nationwide Foreign:
Product codes and lots
syngo.CT software version VA48A_SP5.5

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2440-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2440-2019Class IITerminatedVoluntary: Firm initiated