Z-2478-2019Class II
cobas c513 Analyzer
Roche Diagnostics GmbH / initiated 2019-07-18 / Terminated / root cause: design
Roche Diagnostics Operations, Inc. began a recall of cobas c513 Analyzer on . FDA classified it as Class II and gives the reason as "Quality issue with high pressure solenoid valves". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2478-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- cobas c513 Analyzer
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Quality issue with high pressure solenoid valves
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- 1. For the cobas 8000 ISE, cobas c 701, and cobas c 702 modules and the cobas c 513 analyzer, stop using these units immediately. Your FES will be in contact with you to replace these affected high pressure solenoid valves immediately. 2. In the event that the high pressure solenoid valves cannot be replaced immediately, for the cobas e 601 module, follow the wash station checks outlined in this Urgent Medical Device Correction (UMDC). Perform these checks once per shift. If these checks show favorable outcomes and quality control results are within range, patient results may be reported. If these checks produce unfavorable outcomes, stop using your cobas e 601 module and call the Roche Support Network Customer Support Center at 1-800-428-2336. 3. Complete the attached fax form and fax or email it according to the instructions on the form. 4. File this Urgent Medical Device Correction (UMDC) for future reference Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
- Quantity in commerce
- 2 units
- Distribution
- State NY MT HI IN MD TX OH KS KY CA FL OR LA MO AK
- Product codes and lots
- Serial # :1731-08, 1722-03
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2476-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2477-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2478-2019 | Class II | Terminated | Voluntary: Firm initiated |