Siemens Atellica CH 930 Analyzer. Siemens Material Number (SMN): 11067000
Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Atellica CH 930 Analyzer. Siemens Material Number (SMN): 11067000 Product Usage: automated, clinical chemistry analyzer to perform in vitro diagnostic... on . FDA classified it as Class II and gives the reason as "Atellica CH 930 Analyzer - Erroneous Indices for Hemolysis and Lipemia in Software versions 1.19.2 and below". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0148-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens Atellica CH 930 Analyzer. Siemens Material Number (SMN): 11067000 Product Usage: automated, clinical chemistry analyzer to perform in vitro diagnostic tests on clinical specimens
- Product code
- JGS Electrode, Ion Specific, Sodium
- Reason for recall
- Atellica CH 930 Analyzer - Erroneous Indices for Hemolysis and Lipemia in Software versions 1.19.2 and below
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens issued an Urgent medical Device Correction, (UMDC ASI19-02.A.US) via Fed Ex and Urgent Field Safety Notification (UFSN ASI19-02.A.OUS) to all affected Customers beginning on 2019-07-30. The UFSN/UMDC explain the behaviors that can occur and instructs customers on actions to be taken for the issues described in the communication. Customers are requested to return the attached effectiveness check form via fax or email within 30 days. Field Service personnel will be sent a Service Knowledge Base describing the issue and instructing them how to deal with customer questions.
- Quantity in commerce
- 1125 units
- Distribution
- Worldwide Distribution - US Nationwide Foreign: Argentina Australia Austria Bangladesh Belgium Brazil Bulgaria Canada Chile Colombia Czech Republic Denmark Ecuador Egypt Finland France Germany Greece Hungary India Indonesia Iran Ireland Israel Italy Japan Kuwait Latvia Malaysia Mexico Netherlands Nigeria Norway P.R. China Pakistan Philippines Poland Portugal Qatar Republic Korea Romania Russian Fed. Saudi Arabia Singapore Slovakia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. United Kingdom Uruguay Vietnam
- Product codes and lots
- Software version: V1.19.2 and below