Giraffe OmniBed Carestation CS1 -
GE Healthcare, LLC began a recall of Giraffe OmniBed Carestation CS1 - Product Usage: The device can be operated as an incubator or as a warmer and can transition from one mode to the other on... on . FDA classified it as Class II and gives the reason as "Fasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0281-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Giraffe OmniBed Carestation CS1 - Product Usage: The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo regulate based on their own physiology.
- Product code
- FMZ Incubator, Neonatal
- Reason for recall
- Fasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- GE Healthcare notified customers on about 07/26/2019 via phone calls and a "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were instructed to immediately stop use of and segregate the affected devices, that GE Healthcare would correct all affected products at no cost, and a GE Healthcare representative will contact them to arrange for the correction. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative. A second customer letter was sent on 08/05/2019 to all customer. The second letter contained the same instructions, but provided an updated list of affected serial numbers.
- Quantity in commerce
- 59 devices (50 US, 9 OUS)
- Distribution
- Worldwide distributions - US Nationwide in the states of CA, CT, FL, ID, IL, KS, MI, NC, NC, NJ, NM, PA, TX, VA. Countries of China, France, Germany, Korea, Republic of China.
- Product codes and lots
- Model Number: 2082844-001; Serial Number (UDI Number): TABY70690 (010084068211686221TABY7069011190610), TABY70692 (010084068211686221TABY7069211190611), TABY70708 (010084068211686221TABY7070811190612), TABY70719 (010084068211686221TABY7071911190613), TABY70709 (010084068211686221TABY7070911190612), TABY70724 (010084068211686221TABY7072411190614), TABY70722 (010084068211686221TABY7072211190614), TABY70718 (010084068211686221TABY7064211190603), TABY70717 (010084068211686221TABY7064411190603), TABY70712 (010084068211686221TABY7064511190603), TABY70678 (010084068211686221TABY7067811190607), TABY70725 (010084068211686221TABY7072511190615), TABY70672 (010084068211686221TABY7067211190607), TABY70686 (010084068211686221TABY7068611190610), TABY70688 (010084068211686221TABY7068811190610), TABY70697 (010084068211686221TABY7069711190611), TABY70726 (010084068211686221TABY7072611190616), TABY70729 (010084068211686221TABY7072911190616), TABY70645 (010084068211686221TABY7064711190604), TABY70647 (010084068211686221TABY7066011190605), TABY70646 (010084068211686221TABY7071211190613), TABY70644 (010084068211686221TABY7071711190613), TABY70642 (010084068211686221TABY7071811190613), TABY70732 (010084068211686221TABY7073211190617), TABY70723 (010084068211686221TABY7072311190614), TABY70747 (010084068211686221TABY7074711190619), TABY70748 (010084068211686221TABY7074811190619), TABY70673 (010084068211686221TABY7067311190607), TABY70676 (010084068211686221TABY7067611190607), TABY70693 (010084068211686221TABY7069311190611), TABY70648 (010084068211686221TABY7064811190604), TABY70675 (010084068211686221TABY7067511190607), TABY70694 (010084068211686221TABY7069411190611), TABY70695 (010084068211686221TABY7069511190611), TABY70730 (010084068211686221TABY7073011190617), TABY70734 (010084068211686221TABY7073411190617), TABY70737 (010084068211686221TABY7073711190618), TABY70677 (010084068211686221TABY7067711190607), TABY70698 (010084068211686221TABY7069811190611), TABY70699 (010084068211686221TABY7069911190611), TABY70703 (010084068211686221TABY7070311190612), TABY70706 (010084068211686221TABY7070611190612), TABY70727 (010084068211686221TABY7072711190616), TABY70728 (010084068211686221TABY7072811190616), TABY70660 (010084068211686221TABY7064611190603), TABY70689 (010084068211686221TABY7068911190610), TABY70705 (010084068211686221TABY7070511190612), TABY70707 (010084068211686221TABY7070711190612), TABY70731 (010084068211686221TABY7073111190617), TABY70733 (010084068211686221TABY7073311190617), TABY70652 (Not Available), TABY70654 (Not Available), TABY70662 (Not Available), TABY70664 (Not Available), TABY70710 (Not Available), TABY70715 (Not Available), TABY70716 (Not Available), TABY70753 (Not Available), TABY70659 (Not Available), TABY70704 (010084068211686221TABY7070411190612), TABY70755 (010084068211686221TABY7075511190620), TABY70638 (Not Available), TABY70649 (Not Available), TABY70656 (Not Available), TABY70661 (Not Available), TABY70657 (Not Available), TABY70650 (Not Available), TABY70667 (Not Available), TABY70668 (Not Available), TABY70653 (Not Available), TABY70663 (Not Available), TABY70670 (Not Available), TABY70665 (Not Available), TABY70669 (Not Available), TABY70671 (Not Available), TABY70714 (Not Available), TABY70658 (Not Available), TABY70655 (Not Available), TABY70666 (Not Available), TABY70651 (Not Available).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0281-2020 | Class II | Ongoing | Voluntary: Firm initiated |