Z-2542-2019Class II

Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) System, labeled as EMBLEMTM S-ICD Pulse Generator

Boston Scientific Corporation / initiated 2019-08-21 / Open, Classified / root cause: design

Boston Scientific Corporation began a recall of Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) System, labeled as EMBLEMTM S-ICD Pulse Generator on . FDA classified it as Class II and gives the reason as "Potential need for device replacement - Elective Replacement Indicator (ERI) or End of Life (EOL) - earlier than expected due to compromised performance of an electrical component causing accelerated battery depletion". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2542-2019
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Boston Scientific Corporation
Product
Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) System, labeled as EMBLEMTM S-ICD Pulse Generator
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Potential need for device replacement - Elective Replacement Indicator (ERI) or End of Life (EOL) - earlier than expected due to compromised performance of an electrical component causing accelerated battery depletion.
Root cause tag
design
FDA root cause
Component design/selection
Action
Boston Scientific began notifying physicians of the recall on about 08/21/2019, via "URGENT: MEDICAL DEVICE ADVISORY" letter. The physicians associated with the affected devices were provide a letter with a list of affected model/serial numbers for patients which they have implanted and/or followed or sent to their center. The letter informed physicians of the potential for accelerated battery depletion on select active worldwide EMBLEM Subcutaneous Implantable Cardioverter Defibrillators (S-ICDs). Physicians were instructed to follow up with affected units implanted in patients by enroll and monitoring patients in LATITUDE to facilitate prompt detection of ERI/EOL during the interval between in-office device checks, perform a device follow-up every 3 months via remote or in-office interrogation, promptly investigate any suspected indication of accelerated depletion and contact Boston Scientific Technical Services for assistance as needed, and append the patients medical record with this letter to maintain awareness of this topic for the remaining service life of the device. Physicians were also instructed to evaluate the potential for life-threatening harm due to accelerated depletion for patients are the greatest risk, and replace the affected device within 21 days of ERI and prophylactically for those devices in high risk patients. Boston Scientific also provided an advisory letter for all EMBLEM physicians in order to reiterate and summarize the battery management recommendations that exist in current EMBLEM S-ICD labeling. Included in the customer notification to each physician was (affected and advisory) is a patient advisory letter, which each physician can chose at their discretion to provide. Questions and comments for US customers can be directed to Technical Services, Phone: 1.800.CARDIAC (227.3422) or Email: [email protected]. In addition, Boston Scientific worked with the Heart Rhythm Society who sent an email August 22, 20
Quantity in commerce
28 devices
Distribution
Nationwide distribution to Alabama, Arizona, Connecticut, Delaware, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Kansas, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina, Tennessee, Texas, Vermont, Virginia, Washington, Wisconsin, Wyoming. International distribution to Argentina, Australia, Austria, Bahamas, Belgium, Bermuda, Brazil, Brunei Darussalam, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, French Polynesia, Germany, Greece, Guam, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Lebanon, , Lithuania, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Poland, Portugal, Puerto Rico, Qatar, Reunion, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan R O C, Thailand, Tunisia, Turkey, U.S. - Virgin Islands, United Arab Emirates, United Kingdom.
Product codes and lots
Model A209, Serial Number (Use Before Date) - 213729 (7/24/2019), 213859 (7/26/2019), 213891 (7/27/2019), 213896 (7/27/2019), 213898 (7/27/2019), 213901 (7/27/2019), 213902 (7/27/2019), 213904 (7/27/2019), 213910 (7/27/2019), 213919 (7/27/2019), 213921 (7/27/2019), 213937 (7/27/2019), 213938 (7/27/2019), 213943 (7/27/2019), 213946 (7/27/2019), 213961 (7/28/2019), 213972 (7/28/2019), 213975 (7/28/2019), 213976 (7/28/2019), 213979 (7/28/2019), 213982 (7/28/2019), 213983 (7/28/2019), 214020 (7/31/2019), 214030 (7/31/2019), 214032 (7/31/2019), 214036 (7/31/2019), 214042 (7/31/2019), 214083 (7/31/2019);

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2542-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2542-2019Class IIOngoingVoluntary: Firm initiated
Z-2543-2019Class IIOngoingVoluntary: Firm initiated