TriMed Dorsal Wrist Hook Plate, Dorsal, 4-Hole, REF WHD-4, contained in a tray, insert, or poly package...
TriMed Inc. began a recall of TriMed Dorsal Wrist Hook Plate, Dorsal, 4-Hole, REF WHD-4, contained in a tray, insert, or poly package containing varying amounts, Nonsterile, Single Use... on . FDA classified it as Class II and gives the reason as "Modifications to the specification were not appropriately qualified through the company's design control process which may result in structural integrity issues of the device". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0118-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Trimed Inc
- Product
- TriMed Dorsal Wrist Hook Plate, Dorsal, 4-Hole, REF WHD-4, contained in a tray, insert, or poly package containing varying amounts, Nonsterile, Single Use Only, Rx. The firm name on the label is TriMed, Santa Clarita, CA.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- Modifications to the specification were not appropriately qualified through the company's design control process which may result in structural integrity issues of the device.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The recalling firm issued letters via email on 8/24/2018 informing the distributor of the medical facility of the issue. The distributor is then responsible for notifying the medical facility to whom the device is invoiced.
- Quantity in commerce
- 5 devices
- Distribution
- Distribution was made to CA, FL, MD, TX, and WA. There was military and foreign distribution was made to Australia. There was no government distribution.
- Product codes and lots
- Lot numbers 28884, 28885, 28886, and 28887
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0118-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY