Artis zee and Artis Q/ Q.zen Interventional Fluoroscopic X-Ray System with patient tables as follows: Systems...
Siemens Medical Solutions USA, Inc began a recall of Artis zee and Artis Q/ Q.zen Interventional Fluoroscopic X-Ray System with patient tables as follows: Systems Material Number Artis zee biplane 10094141; Artis... on . FDA classified it as Class II and gives the reason as "Artis zee/Q systems patient tables may potentially have cracks in the table mainframe, and cause mechanical detachment of the upper part of the patient bed and lead to hazardous situations for patients and medical...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2367-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Artis zee and Artis Q/ Q.zen Interventional Fluoroscopic X-Ray System with patient tables as follows: Systems Material Number Artis zee biplane 10094141; Artis Q floor 10848280; Artis Q biplane 10848282; Artis Q.zen floor 10848353; Artis Q.zen biplane 10848355; Artis zee ceiling 10094137; Artis Q ceiling 10848281; Artis Q zeego 10848283 ArtisQ.zen ceiling 10848354 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. The Artis family can also support the acquisition of position triggered imaging for spatial data synthesis.
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Artis zee/Q systems patient tables may potentially have cracks in the table mainframe, and cause mechanical detachment of the upper part of the patient bed and lead to hazardous situations for patients and medical personnel due to falling parts
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice (AX011/19/S) on 8/1/19 a . The letter identifies the reason for recall, health risk and action to take: Siemens Service organization will inspect the table mainframe for possible cracks and will contact you shortly to arrange a date to perform this corrective action. Contact the service organization for an earlier appointment at 1-800-888-7436.
- Quantity in commerce
- US 17
- Distribution
- US Nationwide Distribution - AL, AR, KS, MI, MN, NC, ND, NY, OH, SC, TX, WI
- Product codes and lots
- Serial Numbers: 123245 103302 121501 148401 109589 148393 111129 148400 103310 148397 155235 121505 109563 121506 131235 121502 105125
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2367-2019 | Class II | Terminated | Voluntary: Firm initiated |