Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0211-2020Class II

Beckman Coulter Uric Acid, REF OSR6098, OSR6198, and OSR6698

Siemens Healthcare Diagnostics Inc. / initiated 2018-08-24 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of Beckman Coulter Uric Acid, REF OSR6098, OSR6198, and OSR6698 on . FDA classified it as Class II and gives the reason as "NAPQI, a metabolite of Acetominophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0211-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Beckman Coulter Uric Acid, REF OSR6098, OSR6198, and OSR6698.
Product code
KNK Acid, Uric, Uricase (Colorimetric)
Reason for recall
NAPQI, a metabolite of Acetominophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The recalling firm issued letters dated 8/21/2018 via email or postal mail if an email address was not available on 8/24/2018 to their customers. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-223-0139 in the United States and Canada. Outside the United States and Canada contact local Beckman Coulter Representative.
Quantity in commerce
19,834 uits
Distribution
Distribution was nationwide. There was government and military distribution. Foreign distribution was made to Canada.
Product codes and lots
All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0211-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0210-2020Class IITerminatedVoluntary: Firm initiated
Z-0211-2020Class IITerminatedVoluntary: Firm initiated